4
Installing an Auxiliary Pressure Control Module
Disconnect the GC
Disconnect the GC
WARNING
Hydrogen gas is flammable and potentially explosive. Before replacing the
module, turn off the hydrogen gas at the source.
1.
Turn off the GC and unplug the power cord. Allow time for all heated
zones to cool.
2.
Disconnect all cables from the GC.
3.
Turn all gases off at their sources. Disconnect the carrier and detector
gas tubing from the back panel of the instrument.
Disconnect gas supply tubing
from these fittings