2 - Installation
16
Location
Refer to the Safety Summary page at the beginning of this manual for safety-related information about
environmental conditions.
WARNING:
Agilent 6814B units weigh 79.5 kg (175 lbs.)
Agilent 6834B/ 6843A units weigh 87.7 kg (193 lbs.)
Obtain adequate help when moving the unit or mounting the unit in the rack.
Bench Operation
The outline diagram in figure 2-1 gives the dimensions of your ac source. The feet may be removed for
rack mounting. Your ac source must be installed in a location that allows sufficient space at the sides and
back of the unit for adequate air circulation. Minimum clearances are 1 inch (25 mm) along the sides. Do
not block the fan exhaust at the rear of the unit.
Rack Mounting
The ac source can be mounted in a standard 19-inch rack panel or cabinet. Rack mounting kits are
available as Option 1CM or 1CP. Installation instructions are included with each rack mounting kit
CAUTION:
Agilent 6814B/ 6834B/ 6843A units require instrument support rails for non-stationary
installations. These are normally ordered with the cabinet and are not included with the
rack mounting kits.
Figure 2-1. Outline Diagram
262.6mm
10.3"
12.7mm
0.5"
425.5mm
16.75"
574.7mm
22.6"
266.7mm
10.5"
19.1mm
0.8"
REAR
TOP
SIDE