Troubleshooting - 3
25
Turn on unit and observe the
display. Unit should display all of
the segments and annunciators,
the address and then after self
test display either an error
message or go to the metering
mode.
Protect
Annunciator
On?
Display comes
on?
A3J2-5 held
low?
Yes
Replace A2
Interface board
No
No
+5V @ A3J2-8?
Troubleshoot A1
+5V Interface Bias
circuit, W6 or W7.
No
Yes
RI?
Replace A3 Front
Panel/Display
board
Error Message?
See Error Message
Table 3-2
Yes
Yes
No
OV?
No
OC?
No
Go to
"Troubleshooting
OV at Turn-on"
(sheet 4)
Yes
Check RI input, A2
Interface board
Yes
Disable OCP and
check for normal
operation
Yes
Replace internal
fuse F300
Go to sheet 2
No
FS?
No
For OT check fan
circuit, thermal
sensor RT301
No
Fuse blown?
Yes
Yes
Go to "FS indicated
but fuse is OK"
(sheet 6)
No
Figure 3-1 Sheet 1. Main Flowchart
Summary of Contents for 6632B
Page 8: ......
Page 27: ...Troubleshooting 3 27 Figure 3 1 Sheet 3 Main Flowchart continued...
Page 49: ...Principles of Operation 4 49 Figure 4 1 A2 A3 Block Diagram...
Page 51: ...Principles of Operation 4 51 Figure 4 2 A1 Block Diagram...
Page 54: ......
Page 57: ...Replaceable Parts 5 57 Figure 5 1 Mechanical Parts ldentification...
Page 75: ...Figure 6 1 A1 Board Component Locations...
Page 76: ...Figure 6 2 A4 and A6 Board Component Locations...
Page 77: ...Figure 6 3 A1 Board schematic sheet 1...
Page 78: ...Figure 6 3 A1 Board schematic sheet 2...
Page 79: ...Figure 6 3 A1 Board schematic sheet 3...