![Agilent Technologies 6631B User Manual Download Page 61](http://html.mh-extra.com/html/agilent-technologies/6631b/6631b_user-manual_2868399061.webp)
Verification and Calibration - B
61
Current Programming and High-Range Measurement Calibration
Action
Display
13.
Connect the appropriate current monitor as shown in figure B-1A.
Connect the DMM (in dc mode) across the current shunt.
14.
Press
Shift Cal
, scroll to CAL CURR, and press
Enter
.
CAL:CURR
15.
Press
Shift Cal
, scroll to CAL LEV, and press
Enter
to select the
first calibration point.
CAL:LEV P1
16.
Press
Shift Cal
and scroll to CAL DATA. Wait for the DMM reading
to stabilize. Then read the DMM and compute the first current value
(DMM reading
÷
shunt resistance). Use the Entry keypad to enter the
first current value.
CAL:DATA 0.00
17.
Press
Shift Cal
, scroll to CAL LEV, use
ä
to scroll to P2 (the
second calibration point), and press
Enter
.
CAL:LEV P2
18.
Press
Shift Cal
and scroll to CAL DATA. Wait for the DMM reading
to stabilize. Then read the DMM and compute the second current
value (DMM reading
÷
shunt resistance). Use the Entry keypad to
enter the second current value.
CAL:DATA 0.00
Negative Current Programming Calibration
Action
Display
19.
Connect only an external supply directly to the dc source. Connect the
+ terminal to the + output and the - terminal to the - output. Set the
external supply to 1 V, 1 A. The dc source should be in CC mode.
20.
Press
Shift Cal
, scroll to CAL CURR NEG, and press
Enter
.
CAL:CURR:NEG
21.
Wait for the dc source to compute the negative current calibration
constant. The display returns to Meter mode when the calculation is
complete. Disconnect the external supply.
Low-Range Current Measurement Calibration
Action
Display
22.
Press
Shift Cal
, scroll to CAL CURR MEAS LOW, and press
Enter
.
CAL:CURR:MEAS:LOW
23.
Connect the UUT as shown in figure B-1B with the appropriate load
resistor for your model. Set the DMM to operate in current mode.
24.
Press
Shift Cal
, scroll to CAL LEV, and press
Enter
to select the
first calibration point.
CAL:LEV P1
25.
Press
Shift Cal
and scroll to CAL DATA. Wait for the DMM reading
to stabilize. Then use the Entry keypad and enter the current reading
displayed on the DMM.
CAL:DATA 0.00