B - Performance, Calibration, and Configuration
142
DC Power Supply
8V @ 5A (for current sink verification/calibration)
25 V source (for DVM verification/calibration)
Agilent 6611C, 6631B
6631C, or 6633B
GPIB Controller
Full GPIB capabilities (only required if you are
calibrating the unit over the GPIB)
HP Series 200/300 or
equivalent
Oscilloscope
Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip
Agilent 54504A or
equivalent
RMS Voltmeter
True RMS
Bandwidth: 20 MHz
Sensitivity: 100
µ
V
Agilent 3400B or
equivalent
Variable-Voltage Transformer
Adjustable to highest rated input voltage range.
Power: 500 VA
RC network (required for stable
operation in H-remote mode)
Capacitor: fixed film 25
µ
F, 50V
Resistor: 0.25
Ω
, 1W
p/n 0160-8387
Measurement Techniques
Test Setup
Figure B-1 shows the setup for the tests. Use load leads of sufficient wire gauge to carry the full output
current. An RC network is needed if the unit is operated in LRemote or HRemote mode (see chapter 3).
Load
-S
-
+
+S
resistor
DC
Ammeter
-
+
Load
-S
-
+
+S
resistor
DC
Ammeter
-
+
+
-
External
DC supply
C.
D.
400 ohm
Electronic
-
+
Load
NOTE: Connector
is removable
DVM, Scope, or
RMS Voltmeter
-
+
-S
-
+
+S
A.
NOTE: Connector
is removable
E.
-
+
External DC
-
+
Supply
DC
Voltmeter
-
+
-S
-
+
+S
DVM connector
Output 1
connector
connect this lead only
when calibrating the unit
(only for
verification)
(for CV tests)
Electronic
-
+
Load
-S
-
+
+S
-
+
B.
Current
monitor
NOTE: Connector
is removable
DVM, Scope, or
RMS Voltmeter
(for CC tests)
(see note)
Note: Replace electronic
load with DC supply for
-3A Current Sink test.
Figure B-1. Verification and Calibration Test Setup
p/n 0699-4707