Agilent Technologies 66102A Service Manual Download Page 2

Page 2 of 5 

 

66103A-08A 

 

 
SPECIAL NOTE: 

When a 6610XA comes in for calibration or repair and the instrument has a serial number with in 
the range shown above the socket may need to be replaced?  
 
To identify if the socket is defective see Figure 2 below for specific details.    

 
SPECIAL REPAIR NOTE! 

(Circuit Reference “U001”).

 

The 16-Bit processor 

p/n1821-1479

 should only be replaced if the package is cracked, broken or the pins are bent or 

broken during the modification. 
Remove the 16-Bit processor 

p/n1821-1479

 from the socket before the 68 pin PLCC socket p/n 1200-1274 is 

replaced.  
The 16-Bit processor package 

p/n1821-1479

 

must be inspected for 

cracked bent or broken pins 

before it is reinstalled 

in the NEW socket.

 

 
It is recommended that an authorized technician or operator that has been trained to solder delicate components and 
multi layer printed circuit board assemblies remove and replace the 68 pin PLCC socket 

p/n 

1200-1274.

 

 

 
The 

PCA-TSTD, PWRMESH p/n 

6610X-61022

 

should only be replaced if the traces on the printed circuit board are 

damaged during the removal of the 

68 Pin PLCC 

socket 

p/n 

1200-1274

 

(Circuit Reference “U001”).  

 
It is very unlikely that the PCA-Tested assembly p/n

 

6610X-61022

 

will need to be replaced if

 

the 68 pin PLCC 

socket 

p/n 

1200-1274

 

is removed and replace with care

.  

 

Situation: 

When the 16-Bit processor 

p/n1821-1474

 

is installed in the 68 pin PLCC 

SOCKE

P/N 1200-1274 

“U001” if there is not a gas tight connection between the processer pins and the socket pins an 
intermittent connection will occur.  
 
The pin connections may not be reliable. Over time the pins may become intermittent which will cause 
the instrument not to communicate or not complete self test.  

 
IMPORTANT  

Simply removing the 16-Bit processor 

p/n 1821-1474 and reinstalling it will not insure a gas tight connection 

or guarantee that the instrument will be reliable.  To insure reliability the socket must be removed and replaced. 

 

 
Figure 1 will help to identify the 

68-PIN PLCC SOCKET location - 

P/N 1200-1274 (Circuit Reference 

“U001”)  
                                                                                                        This picture shows the required tool 
                                                                                                         used to remove the processor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  

                  

 

Figure 1 

 

Summary of Contents for 66102A

Page 1: ... E N O T E Supersedes 66103A 08 66103A DC Power Supply Serial Numbers Serial Numbers MY40005783 MY40006268 The instrument will not communicate or will not complete self test Parts Required P N Description Qty 1200 1274 SOCKET IC 68 CONT PLCC 1 1821 1479 IC 16 Bit MCU 16 MHZ CMOS 68PLCC 1 6610X 61022 PCA TSTD PWRMESH IMPORTANT Consult the Service Manual for the 6610XA Power Modules p n 5959 3364 Th...

Page 2: ...ESH p n 6610X 61022 should only be replaced if the traces on the printed circuit board are damaged during the removal of the 68 Pin PLCC socket p n 1200 1274 Circuit Reference U001 It is very unlikely that the PCA Tested assembly p n 6610X 61022 will need to be replaced if the 68 pin PLCC socket p n 1200 1274 is removed and replace with care Situation When the 16 Bit processor p n1821 1474 is inst...

Page 3: ...viewed these markings should be used to identify a Good or Defective socket Socket A is a sample of a GOOD AMP socket Socket B is a sample of a GOOD Tyco Augit socket Socket C is a sample of a GOOD FCI socket Socket D is a sample of a DEFECTIVE Tyco socket The sockets above labeled A B or C are good and should NOT be replaced The socket above labeled D is considered defective should replaced The i...

Page 4: ...nt was tested by comparing the mating and un mating forces on the older and newer designs The test results for sockets with the increased spring height newer design showed an increase in the mean mating force of 18 9 and an increase in the mean un mating force of 23 4 The increase in the un mating force test results between the old design and the new design indicate that a greater normal force is ...

Page 5: ...5 66106A MY40000389 MY40000589 Testing Calibration NOTE When this modification is complete the instrument should be turned on using the 66000A mainframe The 6610XA should complete self test The instrument should also be programmed for voltage and current Complete performance testing is NOT required The instrument should output the programmed voltage and current and should pass the self test Changi...

Reviews: