![Agilent Technologies 5977B Series Troubleshooting And Maintenance Manual Download Page 224](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608224.webp)
224
5977B Series MSD Troubleshooting and Maintenance Manual
8
Electronics
Interfacing to External Devices
Remote control processor
The remote control processor on the LAN/MS control card synchronizes
start-run signals with GCs and other devices. The functions of the remote
control processor are extended to the remote start (Remote) connector
(
) on the back panel of the MSD. The remote start cable connects the
GC and the MSD.
Remote start signals
It is often necessary to communicate with external devices (for example, a
purge-and-trap) during a run. Typically, these communications are requests to
send a system-ready signal. They also include:
•
Receive a start run signal from an external device
•
Program the timing of events during a run
System ready
When interfacing to an external device, it is often desirable to send a
system-ready signal to the device. In the case of a multisample Tekmar
purge-and-trap, each sample is purged onto a trap where it waits for a ready
Figure 54
Remote start connector
Start
Ground
Ready
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......