![Agilent Technologies 5977B Series Troubleshooting And Maintenance Manual Download Page 150](http://html.mh-extra.com/html/agilent-technologies/5977b-series/5977b-series_troubleshooting-and-maintenance-manual_2867608150.webp)
150
5977B Series MSD Troubleshooting and Maintenance Manual
5
CI Maintenance
To Refill the CI Calibrant Vial
Materials needed
•
PFDTD calibrant (8500-8510)
Procedure
1
Set the reagent gas flow to
Gas Off
.
2
Vent the MSD.
3
Remove the capillary column from the GC/MSD interface.
4
Pull the MSD away from the GC to expose the calibration vial and valve.
See
“To Separate the MSD from the 7890 GC”
5
Loosen the collar holding the calibration vial in place. Do not remove the
collar.
6
Remove the calibrant vial. See
7
Fill the vial no higher than the bottom of the internal tube with fresh
PFDTD calibrant (8500-8510).
8
Replace the vial and tighten the collar.
9
Reposition the MSD next to the GC. See
“To Reconnect the MSD to the 7890
10
Reinstall the capillary column.
11
Pump down the MSD.
12
Purge the calibration valve. See
“To Purge the Calibration Valves”
C A U T I O N
Do
not
rinse the vial with any solvents.
Never
expose the inside of the vial to
chlorinated solvents or isopropyl alcohol or water — this will result in severe loss of CI
sensitivity.
C A U T I O N
After removing the calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in severe contamination of the ion source and damage to the filament and
electron multiplier.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......