General Maintenance
4
5977B Series MSD Troubleshooting and Maintenance Manual
143
To Adjust the Quad Frequency
Materials needed
•
Screwdriver, flat-blade, large (8730-0002)
Procedure
1
Ensure the MSD is at thermal equilibrium. It takes at least 2 hours
after
all
heated zones have reached their setpoints for the MSD to reach thermal
equilibrium.
2
Open the analyzer cover.
3
Ensure the RF cover on the side board is secure and no screws are missing.
A loose RF cover or missing screw can
significantly
affect coil adjustment.
4
In the
Tune and Vacuum Control
view, select
Optimize Quadrupole Frequency
from
the Execute menu
.
5
Enter an
m/z
value of 100.
6
Slowly turn the quad frequency adjustment screws to minimize the voltage
displayed. (See
Turn the adjustment screws alternately. Turn each screw only a little bit at
a time. Keep the screws at
equal
extension.
7
When the voltage is minimized, click
Stop
.
WA R N I N G
Do not remove any other covers. Dangerous voltages are present under these
covers.
C A U T I O N
Do not use a coin to adjust the screws. If you drop it, it could fall into the electronics
fan and cause significant damage.
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......