144
5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
8
Repeat steps 4 through 7 for
m/z
650.
9
Exit the Set Optimize Quadrupole Frequency program.
10
Select
MS OFF
from the Execute menu.
11
Close the analyzer cover.
12
Tune the MSD.
Figure 22
Adjusting the quad frequency
RF cover
Quad frequency
adjustment screws