5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
To Reconnect the MSD to the 9000 GC
This procedure starts with both instruments shut down and at room
temperature.
Procedure
1
Position the MSD so the end of the GC/MSD interface is near the GC. (See
2
Tighten the thumb screw at the top of the interface heater clamp. If the
thumb screw is loose when reconnecting the GC/MSD, it will be difficult to
retighten when installing the 9000 GC/MSD Tail.
3
Connect the GC/MSD interface heater cables and LVDS cable.
4
Slide the MSD against the GC with the transfer line entering the GC side
opening, and the metal brackets entering their slots in the base of the GC.
Be careful not to damage the GC/MSD interface as it passes into the GC.
5
Open the GC front door.
6
Slide the MS forward until the GC/MSD interface lightly contacts the bus.
(See
.)
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......