5977B Series MSD Troubleshooting and Maintenance Manual
4
General Maintenance
4
The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
Move it as needed to provide slack in the tubing and cables.
5
Slide the MSD backwards, and then away from the GC until you have
access to the GC/MSD cables. (See
6
Disconnect the GC/MSD interface heater cables, and the LVDS cable.
Disconnecting the interface heater cables with the GC on can cause a fault
condition.
7
Continue to move the MSD until you have access to the part requiring
maintenance.
Figure 6
GC/MSD interface heater cables and LVDS cable
Interface cables
LVDS cable
Summary of Contents for 5977B Series
Page 1: ...Agilent Technologies Agilent 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 8: ...8 5977B Series MSD Troubleshooting and Maintenance Manual ...
Page 33: ...Introduction 1 5977B Series MSD Troubleshooting and Maintenance Manual 33 ...
Page 34: ...34 5977B Series MSD Troubleshooting and Maintenance Manual 1 Introduction ...
Page 88: ...88 3 CI Troubleshooting ...
Page 152: ...152 5977B Series MSD Troubleshooting and Maintenance Manual 5 CI Maintenance ...
Page 180: ...180 5977B Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System ...
Page 226: ...226 5977B Series MSD Troubleshooting and Maintenance Manual 8 Electronics ...
Page 241: ...Parts 9 5977B Series MSD Troubleshooting and Maintenance Manual 241 EI ion source ...
Page 256: ...256 5977B Series MSD Troubleshooting and Maintenance Manual 9 Parts ...
Page 257: ......