5
General Maintenance
5975/5977 Series MSD Operating Manual
157
To Replace a Filament in an EI XTR, SS, or
Inert Source
Materials needed
•
Filament assembly (G7005-60061)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5-mm (8710-1570)
Procedure
1
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
3
Remove the ion source. (See
“To Remove the EI XTR, SS, or Inert Source”
4
Remove the gold plated screw and washer for the filaments. (See
.)
WARNING
The analyzer operates at high temperatures. Do not touch any part until you are
sure it is cool.
Figure 49.
Changing the filament
Gold plated screws
and washers
Gold plated screws
and washers