5
General Maintenance
5975/5977 Series MSD Operating Manual
137
To Install the EI HES Filament
Materials needed
•
Filament assembly, high efficiency dual filament (G7001-60061)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
T6 Torx driver, (8710-2548)
Procedure
1
Remove the old filament. (See
“To Remove the EI HES filaments”
2
Insert the 3 pins on the dual filament through the rear of the filament block.
(See
3
Place the filament block in the source mount.
4
Use a T6 Torx screwdriver to tighten the screw securing the filament block to
the source mount.
5
Reinstall the ion source. (See
6
Close the analyzer chamber. (See
“To Close the Analyzer Chamber”
7
Pumpdown the MSD. (See
8
Autotune the MSD.