To Remove the EI HES filaments
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Cloths, clean (05980-60051)
•
T6 Torx driver, (8710-2548)
Procedure
1
Vent the MS. (See
2
Open the analyzer chamber. (See
“To Open the Analyzer Chamber”
3
Remove the ion source and place it on a clean cloth on your work surface.
(See
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
WARNING
The analyzer, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.