Service Disk Instructions
To reset the attenuator contact counter for the 54845B/46B
1
Turn off the Infiniium oscilloscope
2
Connect a mouse and a keyboard to the oscilloscope
3
Insert the service disk in the floppy drive and turn on the oscilloscope.
4
The oscilloscope will boot into the Windows® operating system.
5
Press Control-Esc on the keyboard to access the Windows® Start menu.
6
Open the Programs group and click on the Infiniium Scope icon.
7
When the oscilloscope application comes up, click on the mouse icon
in the upper right corner of the screen.
8
Click on the Utility pull down menu, click on Selftest. In the Selftest
screen click on Select Service Extensions.
9
Click on the Selftest pull down menu, scroll down and select Atten Set.
Then select the Start Test button.