![Agilent Technologies 54701A User'S And Service Manual Download Page 42](http://html.mh-extra.com/html/agilent-technologies/54701a/54701a_users-and-service-manual_2867914042.webp)
42
Chapter 3: Service
Calibration Testing Procedures
To test input resistance
This test checks the input resistance of the active probe.
Specification: 100 k
Ω
1%
1
Connect the DMM between the probe tip and the ground shell at the
front of the probe.
2
Set up the DMM to measure resistance.
The resistance should read 100 k
Ω
1K
Ω
.
3
Record the reading in the “Calibration Test Record” on page 47.
Equipment Required
Equipment
Critical Specification
Recommended
Model/Part
Digital Multimeter
Resistance 0.1%
3458A
If the test fails
Go to the "Troubleshooting and Repair" section in this chapter.
Summary of Contents for 54701A
Page 7: ...7 1 Operating the Probe ...
Page 21: ...21 2 Probing Considerations ...
Page 33: ...33 3 Service ...
Page 58: ...58 Chapter 3 Service Troubleshooting and Repair ...