Installation
30
Agilent 5100 and 5110 ICP-OES User’s Guide
For instructions on how to install your ICP Expert software, refer
to the documentation provided with your software:
ICP Expert Software Installation Instructions
ICP Expert Software Installation Instructions for 21 CFR Part 11
Environments
Once you have installed the appropriate software, connect the
Ethernet cable, set the instrument IP address if necessary, and then
perform the Detector Calibration and Wavelength Calibration.
Connecting ICP Expert software to the ICP-OES
To connect the software and instrument:
1
Start the ICP Expert software by double-clicking on the desktop
icon.
2
Click
Instrument
.
3
Click
Connect
.
4
Enter the IP address of the instrument or select an existing
instrument.
5
Click
Connect
.
6
Click
Close
.
Performing a Detector and Instrument Calibration
To perform a detector and wavelength calibration:
1
Turn on the external exhaust.
2
Start the ICP Expert software and open the Instrument Window.
3
Connect the software to the instrument, if needed.
a
Click
Instrument
in the ICP Expert toolbar.
b
Click
Connect
.
c
Select the instrument from the list and the click
Connect
.
4
Click
Calibrate
in the
‘
Detector’ section.