6-26 Agilent 4155C/4156C VXIplug&play Driver User’s Guide, Edition 4
Programming Examples for VEE Users
Multi-Channel Sweep Measurements
Multi-Channel Sweep Measurements
To make multi-channel sweep measurements, use the following functions.
Table 6-10
Functions for Multi-Channel Sweep Measurements
A program example is shown in Figure 6-22 on page 6-27. This program measures
bipolar transistor Ic, Ib-Vbe characteristics. The example uses the User Object of the
Agilent VEE. See Figure 6-23 on page 6-28 and Figure 6-24 on page 6-29. The
measurement setup is shown in Figure 6-21.
Figure 6-21
Device Connection and Source Setup for Example Program
Description
Function
Parameters
Output Switch Setup
hp4156b_setSwitch
channel,state
Output Filter Setup
hp4156b_setFilter
channel,state
Integration Time Setup
hp4156b_setInteg
table,time,average
Forces dc bias
hp4156b_force
channel,mode,range,value,compliance,polarity
Sweep Source Setup
hp4156b_setIv
channel,mode,range,start,stop,point,hold,delay,
s_delay,comp,p_comp
Executes measurement
hp4156b_sweepMiv
channel[ ],mode[ ],range[ ],point,source[ ],
value[ ], status[ ]
Disables output
hp4156b_zeroOutput
channel
A
Collector
Base
Emitter
channel: SMU1
mode: V output
value: 0V
range: 2V
comp: 0.1A
polarity: AUTO
channel: SMU2
mode: V output
value: 0V
range: 2V
comp: 0.1A
polarity: AUTO
channel: SMU3
mode: V output
range: 2V
Start: Vstart
Stop: Vstop
point: MeasPoint
hold: 0sec
delay: 0sec
s_delay: 0sec
comp: 0.1A
p_comp: 1W
channel: SMU1
mode: I measure
value: Ic
range: AUTO
status: IcStatus
A
channel: SMU2
mode: I measure
value: Ib
range: AUTO
status: IbStatus
Summary of Contents for 4155C
Page 3: ......
Page 13: ...1 Installation...
Page 19: ...2 Driver Function Reference...
Page 89: ...3 Programming Examples for Visual Basic Users...
Page 147: ...4 Programming Examples for Visual Basic NET Users...
Page 183: ...5 Programming Examples for C Users...
Page 225: ...6 Programming Examples for VEE Users...