Chart Output (Analog Out) Commands
See also "Chart Output", starting on page 98 in chapter 3.
OUTPut
[:STATe] {OFF | ON}
[:STATe]?
These commands enable or disable the chart output (analog output) and
query the output state. When disabled, the output is held at 0 V.
OUTPut
:REFerence:OFFset {<
number> | MIN | MAX}
:REFerence:OFFset?
:REFerence:NULL
These commands set or query the offset value. The offset can be set in the
range of -1.0E9 to 1.0E9. MIN sets –1.0E9 V and MAX sets 1.0E9 V.
The
:NULL
form of this command makes the next measurement value the
null value.
OUTPut
:REFerence:SPAN [{<
number> | MIN | MAX}]
:REFerence:SPAN? [{MIN | MAX}]
:REFerence:GAIN?
These commands set or query the span value. The span can be set in the
range of 5E-9 to 1.0E9. MIN sets 5E-9 and MAX sets 1.0E9. The
:GAIN?
query returns the reciprocal of the span value.
OUTPut
:REFerence:VALue {<
value> | MIN | MAX}]
This command allows you to set a fixed value output. The value can be
set in the range of
±
3.00 V. MIN sets -3.00 V and MAX sets +3.00 V.
Chapter 4 Remote Interface Reference
Chart Output (Analog Out) Commands
156
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......