Temperature Measurement Commands
See "Temperature Measurement Configuration" on page 71 in Chapter 3.
•
The meter stores the last settings used for temperature measurements
in non-volatile memory and uses these values as the default for future
measurements. You can, therefore, make temperature measurements
without having to set the transducer type or reference each time.
•
A
range
setting is not meaningful for temperature measurements.
The
range
parameter is included in the commands to allow you to set
the resolution. For temperature measurements, set the
range
to 1.
Other numbers can be entered and are ignored. The range is reported
as 1 in response to a query.
MEASure:TEMPerature?
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[, 1, {<
resolution> | MIN | MAX | DEF}]
{THER | DEF},DEF[, 1, {<
resolution> | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[, 1, {<
resolution> | MIN | MAX | DEF}]
These commands preset and configure the meter for temperature
measurements using the specified probe type and resolution. The reading
is sent to the output buffer. Set the measurement probe type in the first
parameter, either TC (thermocouple), THER (Thermistor), or FRTD
(four-wire RTD). The DEF (default) option sets the last probe type used.
Set the thermocouple or RTD type in the second parameter. Set to DEF
for thermistors. The optional parameter,
1
, is a place holder to allow you
to set the resolution.
CONFigure:TEMPerature
{TC | DEF},{B | E | J | K | N | R | S | T | DEF}[, 1, {<
resolution> | MIN | MAX | DEF}]
{THER | DEF},DEF[, 1, {<
resolution> | MIN | MAX | DEF}]
{FRTD | DEF},{85 | 91 | DEF}[, 1, {<
resolution> | MIN | MAX | DEF}]
These commands preset and configure the meter for temperature
measurements using the specified probe type and resolution. This
command
does not
initiate the measurement. Set the measurement probe
type in the first parameter, either TC (thermocouple), THER
(Thermistor), or FRTD (four-wire RTD). The DEF (default) option sets
the last probe type used. Set the thermocouple type in the second
parameter. Set to DEF for thermistors. The optional parameter,
1
, is a
place holder to allow you to set the resolution.
Chapter 4 Remote Interface Reference
Temperature Measurement Commands
142
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......