Temperature Measurement Configuration
This section contains information to help you configure the meter for
making temperature measurements. To measure temperature, you will
need a temperature transducer. Transducer descriptions and specific
comments about their usage are given on page 263. The meter supports
three general categories of transducer:
•
RTD (Resistive Temperature Device)
•
THERM (Thermistor)
•
T/C (Thermocouple)
The meter supports specific types of transducers in each category:
RTD’s
R
0
= 4.9
Ω
to 2.1 k
Ω
Thermistor
Thermocouples
α
= 0.00385 (DIN/IEC 751)
α
= 0.00391
5k
Ω
B, E, J, K, N, R, S, T
Once you set the measurement units, probe type, and transducer type,
the setting is stored in
non-volatile
memory. This makes the last probe
type used the default probe type.
Measurement Units
•
The meter can report temperature in
°
C (Celsius),
°
F (Fahrenheit),
or K (Kelvins).
•
The measurement units set are stored in
non-volatile
memory.
•
Front Panel Operation:
Set the units under the
2: UNITS
command
in the TEMPerature menu.
•
Remote operation:
Set the units using the following command:
UNIT:TEMPerature {C|F|K}
3
Chapter 3 Features and Functions
Temperature Measurement Configuration
71
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......