Using the
range
and
resolution
Parameters
With the
MEASure?
and
CONFigure
commands, you can select the
measurement function, range, and resolution all in one command. Use
the
range
parameter to specify a fixed range larger than the expected
value of the input signal. You can set the range parameter to AUTO to
set autoranging.
Use the
resolution
parameter to specify the desired resolution for the
measurement. Setting the resolution sets the integration time of the
meter. The resolution parameter can be set to any number but the
resulting integration time (NPLC) is one of the following:
<
resolution>
Parameter
Integration Time
(NPLC)
< 0.0001 x Range
0.02
< 0.00001 x Range
.2
< 0.000003 x Range
1
< 0.0000022 x Range
2
< 0.000001 x Range
10
< 0.0000008 x Range
20
< 0.0000003 x Range
100
< 0.00000022 x Range
200
Specify the resolution in the same units as the measurement function,
not in number of digits
. For example, for volts measurements, specify the
resolution in volts. For resistance, specify the resolution in ohms.
You must specify a range to use the resolution parameter.
4
Chapter 4 Remote Interface Reference
Simplified Programming Sequence
129
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......