160
WAVeform Commands
YINCrement?
YINCrement?
Query
:WAVeform:YINCrement?
The :WAVeform:YINCrement? query returns the y-increment voltage value for
the currently specified source.
This voltage value is the voltage difference between two adjacent waveform data
digital codes.
Adjacent digital codes are codes that differ by one least significant bit. For
example, the digital codes 24680 and 24681 vary by one least significant bit.
Returned Format
<real_value><NL>
<real_value> ::= A real number in exponential format.