Chapter 4 Remote Interface Reference
The SCPI Status Registers
107
4
To Determine When a Command Sequence is Completed
1
Send a device clear message to clear the power supply’s output buffer
(e.g.,
CLEAR 705
).
2
Clear the event registers with the
*CLS
(clear status) command.
3
Enable the “operation complete” bit (bit 0) in the Standard Event
register by executing the
*ESE 1
command.
4
Send the
*OPC?
(operation complete query) command and enter the
result to ensure synchronization.
5
Execute your command string to program the desired configuration,
and then execute the
*OPC
(operation complete) command as the last
command. When the command sequence is completed, the “operation
complete” bit (bit 0) is set in the Standard Event register.
6
Use a serial poll to check to see when bit 5 (standard event) is set in the
Status Byte summary register. You could also configure the power
supply for an SRQ interrupt by sending
*SRE 32
(Status Byte enable
register, bit 5).
Using *OPC to Signal When Data is in the Output Buffer
Generally, it is best to use the “operation complete” bit (bit 0) in the Standard
Event register to signal when a command sequence is completed. This bit is
set in the register after an
*OPC
command has been executed. If you send
*OPC
after a command which loads a message in the power supply’s output
buffer (query data), you can use the “operation complete” bit to determine
when the message is available. However, if too many messages are generated
before the
*OPC
command executes (sequentially), the output buffer will fill
and the power supply will stop processing commands.
Summary of Contents for 2110-0457
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 26: ...2 General Information...
Page 38: ...3 Front Panel Operation and Features...
Page 70: ...4 Remote Interface Reference...
Page 122: ...5 Error Messages...
Page 133: ...Chapter 5 Error Messages Calibration Errors 132...
Page 134: ...6 Application Programs...
Page 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Page 146: ...7 Tutorial...
Page 157: ...Chapter 7 Tutorial Remote Programming 156...
Page 158: ...8 Specifications...
Page 166: ...Appendix Service Information...
Page 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......
Page 209: ...208 Index Index...