54
16533p01
Single-Card Module
PROTECTED
UNPROTECTED
If you calibrate this module without unprotecting the memory, the new calibration settings will
not be saved when the system is shut down. The system will default to the previous settings.
The new calibration settings would be effective for the current active session only.
Remove the module from the
mainframe and set the
PROTECTED / UNPROTECTED
switch to UNPROTECTED.
2
16700A/B
16702A/B
Power down the mainframe.
1
16533/34A Calibration
16600A Series