24
Agilent 16442A User’s Guide, Edition 3
User's Guide
Specifications
Specifications
The "supplemental information" and "typical" entries, in the following
specifications are not warranted, but provide useful information about the functions
and performance of the instruments.
The following specifications data is specified at 23 ± 5
°
C and 50 % relative
humidity.
Functions
Agilent 16442A lets you test various shaped DUTs. The 16442A has:
•
6 SMU channels (when using non-Kelvin connections).
3 SMU channels (when using Kelvin connections).
•
2 VSU channels.
•
2 VMU channels.
•
2 PGU channels.
•
1 GNDU channel.
•
interlock (Intlk) 6-pin connector.
Voltage and current range
Channel
Maximum Voltage
Maximum Current
SMU
200 V
1 A
VSU
40 V
100 mA
VMU
40 V
PGU
40 V
200 mA (AC peak)
GNDU
40 V
1.6 A