Chapter 1
13
Operation
Typical Characteristics
Typical Characteristics
NOTE
f is the measurement frequency in MHz. The incremental error is available at the frequency
range 1 MHz to 15 MHz, and incremental error applies after the completion of
ZERO-OPEN/SHORT compensation.
Figure 1-2
Incremental Error
Table 1-3
Typical Characteristics
Model
Applicable Measurement Range
Incremental error (1 MHz
≤
f )
Parameter Value
Measurement
Frequency
Parameter reading
error (%)
Offset value for D
4192A
Full range
Full range
±
2
×
(f/10)
2
%
*1
±
0.02
×
(f/10)
2
%
4194A
below 15 MHz
4274A
Full range
-
-
4275A
±
2
×
(f/10)
2
%
±
0.02
×
(f/10)
2
%
4276A
-
-
4277A
*2
±
0.02%
±
0.0002%
4278A
*1.There are some cases where the actual incremental error exceeds this limit when this fixture is used with an 4192A
or 4275A.
*2.The maximum measurement frequency of the 4277A and 4278A is 1 MHz, so the incremental error is available
only for 1 MHz measurements.