2D-LC User Guide
49
3
Compatibility Matrix
Supported Firmware
Supported Firmware
Use the firmware, that is available in the Agilent 2D-LC Software USB flash drive
in folder Firmware.
Agilent 2D-LC Software has been tested with following firmware revisions:
Table 3
Supported Firmware
Device
Firmware
Agilent 1100 Series, 1200 Series, and 1200 Infinity
A.07.34
Agilent 1200 Series, 1200 Infinity, and 1120 Compact LC
B.07.34
Agilent 1200 Infinity Hosted Modules
C.07.34
Agilent 1290 Infinity II Modules
D.07.34
Agilent Q-TOF, e.g., G6546A
27.809
1
1
Check the firmware instrument revision codes and firmware versions required for MH Acquisition for TOF
/ Q-TOF version. TOF and Q-TOF firmware revisions are M.XXX with M = Model identifier and XXX = FW re-
vision.
NOTE
•
Agilent releases LC firmware updates for so-called “firmware sets.”
•
All Agilent LC instrument firmware sets have been designed and tested to be
truly and strictly backwards compatible for the installed software base (CDS).
•
The latest module firmware contained in each set is fully compatible and
interoperable with all other module firmware of the same set.
•
Agilent always recommends using the latest module firmware revision of a
firmware set to avoid interoperatibility issues.
•
Generally Agilent always recommends keeping the LC instrument firmware
current.
•
Do not mix firmware revisions between different sets. Agilent does not
guarantee operation of mixed firmware revisions from older or newer sets.
NOTE
Firmware can be found and is available under the following link:
https://www.agilent.com/en-us/firmwareDownload?whid=69761
.
Alternatively, see