10
Troubleshooting and Diagnostics
Lab Advisor Service
&
Diagnostic
2D-LC User Guide
292
Replace the Module Firmware
When
The installation of newer firmware might be necessary
•
if a newer version solves problems of older versions or
•
to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
•
to keep all systems on the same (validated) revision or
•
if a new module with newer firmware is added to a system or
•
if third party control software requires a special version.
Tools required
Description
Agilent Lab Advisor software
Parts required
#
Description
1
Firmware, tools and documentation from Agilent web site
Preparations
Read the following:
•
Documentation provided with the Agilent Lab Advisor online help
•
2D-LC Manual
Procedure to follow:
•
Close the current Acquisition client window
•
Close instrument connection from OpenLab Control Panel
NOTE
Do not mix firmware files from different firmware sets.