![Agilent Technologies 1290 Infinity II 2D-LC Solution User Manual Download Page 127](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575127.webp)
2D-LC User Guide
127
6
Method Parameters
Set the 2D-LC Method parameters
Set the 2D-LC Method parameters
2D LC Operation Mode
Setting the mode has the following consequences:
Heart-Cutting (LC-LC)
The Heart-Cutting mode covers two 2D-LC applications Heart Cutting (LC-LC)
and High-Resolution Sampling (
HiRes
). Once you have selected the Heart-Cutting
mode, you can later define in the software whether you want to use one or the
other mode or even both together.
In Heart-Cutting modus, a relevant volume of
1
D is cut off and injected onto the
2
D column using the
2
D pump. A peak trigger or a time window defines the
volume to be injected on the
2
D column. When heart-cutting starts, a loop is filled
with the peak of interest. Then the injection on the
2
D starts running the gradient
of the
2
D pump.
For details Setting this mode, see
In contrast to Heart-Cutting, which uses the continuous flowthrough principle, in
high-resolution sampling (
HiRes
) the Multiple Heart-Cutting (
MHC
) valve is
switched before and after parking the peak.
When setting up the experiments, keep the following general considerations in
mind:
•
Each loop for consecutive snips stores the same sample volume.
•
First and last loop cannot be used for parking.
•
Solvent transfer from
1
D to
2
D can be reduced.
•
Cut number 5 cannot be parked entirely in the Sample Loop. Otherwise cut 6
would got partially to the transfer capillary and would therefore be lost or spoil
cut 5. Cut 5 stays partially in the transfer line and is immediately being
analyzed in
2
D.
•
For parking cut 6 in the Sample Loop, the cut first needs to be moved from the
2D-LC Valve to the deck valve. This new volume must be defined in the
configuration of the 2D-LC system.
For details Setting this mode, see