HPLC-Chip Cube Interface User’s Guide
49
Installation
4
Chip Cube Hardware Installation
Chip Cube Hardware Installation
1
Open the front door of the Chip Cube.
2
Remove the protective foam pads.
3
Turn the chip feeder / front door transportation lock into the unlock
position.
4
Remove the strap that holds the stages assembly down.
5
Remove the two plastic wedges that fix the stages assembly horizontally.
6
Remove the transport lock (plastic plate) from the stages assembly.
7
Mount the Chip Cube onto the Chip Cube Interface Assembly. The holding
bolts fit into the slotted holes of the Chip Cube Interface Assembly (spray
chamber).
8
Open the Chip Cube front door and connect the Chip Cube to any free CAN
port of the Agilent 1260 Infinity stack using the supplied 1 m CAN cable
(see
9
Connect the power cable to the Chip Cube.
10
Install the capillaries (see
11
Place the open end of the 100 µm waste capillary (black) into a small
beaker.
12
Use the supplied union to connect the open end of the 75 µm capillary to
the 100 µL glass syringe from the Chip Cube Interface Assembly accessory
kit. Install the syringe onto an existing infusion pump (to be provided by
the customer).
13
Load the Protein ID Chip.