![Agilent Technologies 1260 Infinity II User Manual Download Page 14](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii/1260-infinity-ii_user-manual_2867537014.webp)
14
Agilent 1260 Infinity II Preparative Autosampler User Manual
1
Introduction
Operating Principle
11
The gripper arm moves out of the needle station and the wash port snaps
back in position.
12
The gripper arm replaces the vial, and returns to the home position.
Simultaneously, the needle lowers into the seat.
13
The injection valve switches to the mainpass position.
Injection Sequence
Before the start of the injection sequence, and during a preparative run, the
injection valve is in the mainpass position (see
on page 14). In this
position, the mobile phase flows through the autosamplers metering device,
sample loop, and needle. This ensures that all parts in contact with sample are
flushed during the run, thus minimizing carry-over.
Figure 2
Mainpass Position - standard position during runs and when the sampler
is idle
9LDO
:DVK
SRUW
1HHGOH
VHDW
6DPSOLQJORRS
6ROYHQW
:DVK
SXPS
0HWHULQJ
GHYLFH