Agilent 1200 Series Autosampler Reference Manual
253
Theory of Operation
8
Switches 1 and 2 define which set of parameters (for example, for GPIB,
RS-232C, and so on) will be changed. Once the change has been completed, the
instrument must be powered up again in order to store the values in the
non-volatile memory.
In the non-volatile memory the parameters are kept, regardless of whether you
turn the instrument off and on again. They will be kept until the same set of
parameters is subsequently changed and power is reset. All other previously
stored configuration settings will still remain in the non-volatile memory.
In this way you can store more than one set of parameters using the same 8-bit
configuration switch twice, for example, for both GPIB and RS-232C.
GPIB Default Addresses
If you just want to change the GPIB address and need a detailed procedure,
refer to the
Installing Your Agilent ChemStation System
handbook.
Default GPIB address is set to the following addresses:
where 0 means that the switch is down and 1 means that the switch is up.
Table 51
Default Addresses for Agilent 1200 Series Modules
Module
Address
Binary Address
Pump
22
0 0 0 1 0 1 1 0
FLD
23
0 0 0 1 0 1 1 1
VWD
24
0 0 0 1 1 0 0 0
Agilent 8453A
25
0 0 0 1 1 1 0 1
DAD / MWD
26
0 0 0 1 1 0 1 0
Column compartment
27
0 0 0 1 1 0 1 1
Autosampler
28
0 0 0 1 1 1 0 0
RID
29
0 0 0 1 1 1 0 1
Summary of Contents for 1200 series
Page 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Page 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Page 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Page 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Page 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Page 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Page 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Page 277: ......