System Operation Checks
System Operation Checks
Use the following procedures and the standard section in this kit to verify the operation of
your calibrated system.
NOTE
If you are using a different analyzer, refer to its documentation for specific
measurement instructions.
8510 System Operation Check
1. Short the standard section and perform an S
11
and an S
22
measurement to verify that
the ripple on the reflection of the shorted waveguide section does not exceed the limits
listed in the tables on
. If the ripple is less than the acceptable value, then the
combination of directivity and source match errors of the calibrated system is within
the factory specified values.
2. Check the S
11
and S
22
of a thru measurement. As in the previous step, the results of
this check contain the directivity error, but, because of the thru connection to port 2,
this measurement also contains errors contributed by other terms, including load
match, the S
11
and S
22
of the standard sections, and transmission tracking. The actual
or absolute return loss of the standard section should not exceed the acceptable values
listed on
3. Make an S
21
and an S
12
measurement. Ideally, the standard section is low loss. These
measurements verify that the combination of source match error and load match error
do not cause excess ripple on the trace. The trace ripple should not exceed the
acceptable values listed in the tables on
4. Record the results of each test in
through
The S
11
and S
22
of a Shorted Standard Section
1. Set the averaging factor to 512.
2. Perform a TRL calibration.
3. Turn on the calibration.
4. Connect the standard section to port 1.
5. Connect the short to terminate the standard section.
6. Connect the load to port 2.
Summary of Contents for 11644A Series
Page 5: ...X P and K 11644A 1 1 1 General Information ...
Page 10: ...1 6 X P and K 11644A General Information Preventive Maintenance ...
Page 11: ...X P and K 11644A 2 1 2 Specifications ...
Page 17: ...X P and K 11644A 3 1 3 User Information ...
Page 24: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition ...
Page 25: ...X P and K 11644A 4 1 4 Use Maintenance and Care of the Devices ...
Page 34: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage ...
Page 35: ...X P and K 11644A 5 1 5 Performance Verification ...
Page 50: ...5 16 X P and K 11644A Performance Verification System Operation Checks ...
Page 51: ...X P and K 11644A 6 1 6 Troubleshooting ...
Page 55: ...X P and K 11644A 7 1 7 Replaceable Parts ...
Page 56: ...7 2 X P and K 11644A Replaceable Parts ...