X, P, and K 11644A
5-11
Performance Verification
System Operation Checks
Table 5-1 Using a TRL Calibration X11644A WR-90 (X-band)
Test
Measured
Value
Acceptable
Value
Pass
(Yes/No)
S
11
, shorted
≤
0.15 dB
p-p
S
22
, shorted
≤
0.15 dB
p-p
S
11
, thru
≤ −
45 dB
S
22
, thru
≤ −
45 dB
S
21
, thru
≤
0.017 dB
p-p
S
12
, thru
≤
0.017 dB
p-p
Table 5-2 Using a TRL Calibration P11644A WR-62 (P-band)
Test
Measured
Value
Acceptable
Value
Pass
(Yes/No)
S
11
, shorted
≤
0.26 dB
p-p
S
22
, shorted
≤
0.26 dB
p-p
S
11
, thru
≤ −
42 dB
S
22
, thru
≤ −
42 dB
S
21
, thru
≤
0.021 dB
p-p
S
12
, thru
≤
0.021 dB
p-p
Table 5-3 Using a TRL Calibration K11644A WR-42 (K-band)
Test
Measured
Value
Acceptable
Value
Pass
(Yes/No)
S
11
, shorted
≤
0.40 dB
p-p
S
22
, shorted
≤
0.40 dB
p-p
S
11
, thru
≤ −
40 dB
S
22
, thru
≤ −
40 dB
S
21
, thru
≤
0.030 dB
p-p
S
12
, thru
≤
0.030 dB
p-p
Summary of Contents for 11644A Series
Page 5: ...X P and K 11644A 1 1 1 General Information ...
Page 10: ...1 6 X P and K 11644A General Information Preventive Maintenance ...
Page 11: ...X P and K 11644A 2 1 2 Specifications ...
Page 17: ...X P and K 11644A 3 1 3 User Information ...
Page 24: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition ...
Page 25: ...X P and K 11644A 4 1 4 Use Maintenance and Care of the Devices ...
Page 34: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage ...
Page 35: ...X P and K 11644A 5 1 5 Performance Verification ...
Page 50: ...5 16 X P and K 11644A Performance Verification System Operation Checks ...
Page 51: ...X P and K 11644A 6 1 6 Troubleshooting ...
Page 55: ...X P and K 11644A 7 1 7 Replaceable Parts ...
Page 56: ...7 2 X P and K 11644A Replaceable Parts ...