12
General Information
General Characteristics
General Characteristics
The following table gives general characteristics for the probe.
Environmental Conditions
Operating
Non-operating
Temperature
0°C to +55 C° (32°F to +131°F)
-40°C to +70°C (-40°F to +158°F)
Humidity
up to 95% relative humidity (non-
condensing) at +40°C (+104°F)
up to 90% relative humidity at +65°C
(+149°F)
Altitude
up to 4,600 meters (15,000 ft)
up to 15,300 meters (50,000 ft)
Vibration
Random vibration 5 to 500 Hz, 10
minutes per axis, 0.3g
rms
.
Random vibration 5 to 500 Hz, 10 min.
per axis, 2.41 g
rms
. Resonant search 5
to 500 Hz swept sine, 1 octave/min.
sweep rate, (0.75g), 5 min. resonant
dwell at 4 resonances per axis.
Power
Requirements
+12 Vdc @ 33 mA max
-12 Vdc @ 16 mA max
(voltages supplied
+6 Vdc @ 80 mA max
from AutoProbe Interface)
-6 Vdc @ 130 mA max
Weight
approximately 0.59 (1.3 lb)
Dimensions
Refer to outline drawings on the next page.
Summary of Contents for 1153A
Page 4: ...5 1 General Information ...
Page 12: ...13 General Information General Characteristics Figure 1 3 1153A Differential Probe Dimensions ...
Page 13: ...15 2 Operating the Probe ...
Page 23: ...25 3 Service ...
Page 34: ...36 Service Replaceable Parts Figure 3 7 1153A Differential Probe Parts ...