38
1141A and 1142A User’s Guide
2
Calibrating the Probe
CMRR Test
CMRR Test
This test checks the CMRR at 1 MHz and 100 MHz.
Specification: . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3000:1 at 1 MHz, 10:1 at 100 MHz
1
Connect the probe power connector to the
PROBE
connection on the
rear of the 1142A Probe Control and Power Module.
2
Set the 1142A front panel switches to Local and Zero offset.
3
Connect the probe output to the oscilloscope channel 1 input.
4
Connect the input of the probe to the test board in the position
shown in the
Figure 18
Signal to +Input
5
Connect the signal generator to the test board.
6
Set the signal generator for 1 MHz at 385 mV
rms
(1 V
p- p
).
7
On the oscilloscope, press
AUTOSCALE
and set the following
parameters.
Table 9
Equipment Required
Equipment Required Critical Specifications
Recommended
Agilent
Model/Part
Oscilloscope
400 MHz bandwidth at 1 mV/div
54830B
Signal Generator
1-100 MHz at = 400 mVrms
8648A
Test Board
No substitute
01141-66504
Cable
Type N (m) 24-inch
11500B
Adapter
Type N (f) to BNC (m)
1250-0077