Calibrating the Probe
100x Attenuator Accuracy Test
100x Attenuator Accuracy Test
NOTE
If the gain test for the probe fails, it will be reflected in the test for the 100x
attenuator adapter. Do not continue until the probe passes the gain test.
1
Disconnect the probe/attenuator from the test board. Remove the
10x attenuator adapter from the probe and connect the 100x
attenuator adapter.
2
Carefully connect the input of the probe/attenuator to the test
board in the position shown in
.
Figure 15
Signal to +Input
CAUTION
Avoid excessive power dissipation in the termination on the test board. Keep the
voltage input at or below ±7 Vdc.
3
Set the dc source output to 7 Vdc.
4
Record the V
in1
measurement from the top DVM in
5
Record the V
out1
measurement from the bottom DVM in
6
Carefully connect the input of the probe/attenuator to the test
board in the position shown in
Figure 16
Signal to –Input