Optimizing the Fluorescence Detector
3
1100 Series FD Reference Manual
67
Changing the Xenon Flash Lamp Frequency
Modes
The lamp flash frequency can be changed into the following modes:
Best sensitivity can be expected with “
no economy
”, see
Table 16
Flash Lamp Modes
Positioning
296 Hz (Standard), 560 V
63 mJoule (18.8 W)
74 Hz (Economy), 560 V
63 mJoule (4.7 W)
Rotation (Multi Ex/Em)
74 Hz (Standard), 950 V
180 mJoule (13.3 W)
74 Hz (Economy), 560 V
63 mJoule (4.7 W)
Figure 28
Xenon Flash Lamp Frequency
Standard=296 Hz
Economy=74 Hz