First Steps with the Fluorescence Detector
2
1100 Series FD Reference Manual
37
Method Development
Fluorescence detectors are used in liquid chromatography when superior
limits of detection and selectivity are required. Thorough method
development, including spectra acquisition, is fundamental to achieve good
results. This chapter describes three different steps that can be taken with the
Agilent 1100 Series fluorescence detector. Table 5 gives an overview of how to
benefit from the operation modes during these steps.
Table 5
Steps for thorough method development
Step 1:
Check system
Step 2:
Optimize limits of detection
and selectivity
Step 3:
Set up routine methods
Fluorescence scan
Find impurities (for example,
in solvents and reagents)
Determine simultaneously
the excitation and emission
spectra of a pure
compound
Signal mode
Perform wavelength
switching
Use for lowest limits of
detection
Spectral
mode/multi-wavelength
detection
Determine Ex/Em spectra for
all separated compounds in a
single run
Activate up to four
wavelength simultaneously
Collect online spectra,
perform library search,
determine peak purity
Deactivate wavelength
switching
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Page 12: ...12 1100 Series FD Reference Manual ...
Page 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Page 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Page 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Page 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Page 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Page 345: ......