Agilent 10400B Series Miniature Passive Oscilloscope Probes
Bandwidth considerations
Bandwidth considerations
The dominant probe limitation to system bandwidth is its input capacitance,
assuming that the high-frequency compensation adjustments have been made.
The displayed bandwidth of any measurement system using an oscilloscope and
probe is determined by four factors: probe input capacitance, source impedance,
source bandwidth, and oscilloscope bandwidth.
Agilent 10400B Series Probes Selection Guide
The Agilent 10437B and Agilent 10442B probes are compatible with
oscilloscopes that have a 50
Ω
input.
Probe compatibility
Model
Number
Type of Probe Bandwidth
Scope/
Probe
System
Division
Ratio
Input R Input C
Scope
Input R
Comp
Range
Length
(in
meters)
10433B
High-
impedance,
passive
300 MHz
10:1
10 M
Ω
10 pF
1 M
Ω
10-16 pF 2
10436B
High-
impedance,
passive
300 MHz
10:1
10 M
Ω
11 pF
1 M
Ω
18-22 pF 2
10437B
50
Ω
Resistive
Divider
1 GHz
1:1
50
Ω
n/a
50
Ω
n/a
2
10439B
High-
impedance,
passive
n/a
1:1
1 M
Ω
n/a
1 M
Ω
n/a
1.5
10440B
High
impedance
passive
300 MHz
100:1
10 M
Ω
2.5 pF
1 M
Ω
6-14 pF
2
10441B
High-
impedance,
passive
500 MHz
10:1
1 M
Ω
9 pF
1 M
Ω
6-9 pF
1.8
10442B
500-
Ω
resistive
divider
1 GHz
10:1
500
Ω
n/a
50
Ω
n/a
1.5
Infiniium
Oscilloscope
Compatible 10400B Series Probe
8000A Series
10433B, 10437B, 10439B, 10440B, 10442B