Cleaning
To clean the exterior of the equipment:
1.
Stop the system
WARNING:
When the equipment is going to be cleaned, be sure to turn
OFF the power of each device, and to unplug the power cord
from the AC outlet. Never use anhydrous or high solvency
alcohols, benzine, thinner or any other flammable cleaning
agent. Otherwise, it may result in fire or electric shock.
2.
Wipe the exterior of the system with a cloth slightly moistened with a
neutral detergent. Some approved disinfectants can be used for cleaning
as well.
CAUTION:
Make sure no liquid gets in the device.
CAUTION:
Clean the equipment with only a little moisture. Do not
spray disinfectants or detergents directly on the equipment.
Do not pour liquid directly on the equipment.
CAUTION:
Liquids ingressing the DR Detector or the battery may cause
malfunction and contamination. Take special care near the
battery bay and near the cable connector on the side of the
DR Detector.
CAUTION:
Do not use abrasive brush and scraper to clean the product.
Note
:
Do not open the equipment for cleaning. No
components inside the device require cleaning by the user.
3.
Start up the system.
Related Links
on page 56
XD 10, XD
⁺
10 | Introduction to the DR Detector |
53
0432A EN 20210426 1355