DD+DIS033.07E
Repair and Service
Troubleshooting
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.3 / 4 of 72
03-2011 Type
5367/100
Agfa Company Confidential
1
General Troubleshooting Procedure
The troubleshooting procedure depends on the symptom of the fault. Refer to Figure 1:
The online diagnosis has checked
all possible reasons for the fault.
The only possible cure:
To replace a certain module
The online diagnosis has detected,
that more than one possible reason
for the fault exists.
The configured number of cycles
for a certain maintenance action
has been reached.
The online diagnosis was not able to
start up.
A sporadic or persistent film jam
occured.
The image sent from the modality
is not printed. It does not appear
in the printer queue.
The image(s) show stripes, dots,
density or contrast faults.
For more than 5 minutes the printer
shows message
"Please wait selftest proceeding".
Refer to
table next page.
See also
Troubleshooting
Guide.
See also
Troubleshooting
Guide.
See also
Troubleshooting
Guide.
Refer to IMPORTANT
note next page.
Refer to IMPORTANT
note next page.
Refer to
Troubleshooting
Guide.
refer to
Troubleshooting
Guide.
Refer to
Troubleshooting
Guide.
Refer to
Troubleshooting
Guide.
Refer to
Troubleshooting
Guide.
See MEDNET, GSO Library: Hardcopy - Dry Imagers -
Drystar Axys - Troubleshooting Guide
*
*
*
*
*
*
*
*
MODXX
Please replace module
The online diagnosis has detected a
fault, which still allows printing, however
The print quality is reduced or
The print quality can be reduced in the
near future or
A defective part requires replacement
in the near future
New Error: Exxxxxx
xxxxxxxxxxxxxxxxxxxxxxxxx
New Warning: Wxxxxxx
xxxxxxxxxxxxxxxxxxxxxxxx
New Maintenance: Mx
xxxxxxxxxxxxxxxxxxxxxxxxx
Error
System Failure
Film Jam!
XXXXXXXXXXXXXXXXXXXX
Modality Name Time
Film X of Y
Z %
536733aa.cdr
Please wait
Selftest Proceeding
*
Figure 1
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......