DD+DIS150.05E
Repair and Service
Machine specific tools, software tools and auxiliary means
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
Drystar 5302
Chapter 3.2/17
2.3.5 Testimage
LinCheck
PURPOSE
Test image to check density reproduction.
USAGE
(1) In the browser pages select menu
'Service Engineer Tools - Enter
Service Command Session - 3
Print Testimage - LinCheck.ZIP'.
(2) The test image as displayed in
Figure 12 will be printed.
536632ga.cdr
Figure
12
EVALUATION
(1) The test image contains a grey
wedge. This grey wedge is
measured during the density
reproduction check.
Refer to chapter 3.6,
Adjustments.
536632gb.cdr
Density measurement area
Figure
13
For check of density reproduction only measure in the density measurement
area, as there also the CDM is measuring.
2005-09-09 (Type
5366/100)
Agfa Company Confidential