DD+DIS149.05E
Maintenance
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
DRYSTAR 5302
Chapter 9 / 9
3.3
Image quality check
(1) In the browser pages select menu
"Service Engineer Tools - Enter
Service Command Session - 3
Print Testimage"
(2) Print following test images on the
largest format available in the film
magazines:
•
STI1.ZIP
•
STI2.ZIP
•
medical_image.ZIP
The test image as displayed
in Figure 4 to Figure 6 will be
printed.
(3) Mark the test films with a
waterproof pen as testfilm before
maintenance
Figure 4
Figure 5
Figure 6
2005-09-09 (Type
5366/100)
Agfa Company Confidential