DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
2.5.4.3
Performing a Film Calibration after the internal Densitometer Calibration
NOTE:
The film calibration after the internal densitometer calibration has to be performed one
time for each tray.
(1) Select menu “Service Engineer
Tools - Enter Service Command
session - 6 Calibration - 1 Film
Calibration”.
(2) Select the 1
st
input tray: The
calibration film as shown in Figure
14 is printed and the measured
Dmax is shown.
INPUT TRAY:
FILM CALIBRATION OK
Dmax = 3.69 (XR310)
Default densitometer
536736az.cdr
Measured Dmax
(3) Repeat the film calibration for the
2
nd
input tray.
Figure 14
2.5.4.4
Performing a Print Head Calibration followed by two Film Calibrations
NOTE:
The print head calibration has to be performed 2 times, if films of width 10” and 14” are
available.
Instructions
(1) Clean the thermal head with a cleaning tissue lightly moistened with Isopropyl -
alcohol. The last cleaning action must be carried out in one way.
(2) In the browser pages - menu 'Service Engineer Tools - Enter Service Command
session' select '6. Calibration - 3. Print Head Calibration'.
(3) Follow the instructions on the browser pages. Refer to Figure 15.
(4) Perform two film calibrations for both trays. Refer to section 2.5.4.1.
(5) Repeat the print head calibration, if films of width 10” and 14” are available.
Edition 1, Revision 1
DRYSTAR 5302
Chapter 3.6 / 24
05-2009 Type
5366/100