Fixed DR Detector (4343R, Pixium RAD 4343) Exposure
Sequence
For using the fixed DR Detector, the temperature difference between
calibration and usage must be within the recommended range of +/-6°C (for
a DR Detector with CsI conversion screen) or +/-10°C (for a DR Detector with
GOS conversion screen). Check the environmental conditions and observe the
warming-up time of the DR Detector.
Topics:
•
•
Modality Settings
Depending on the DR Detector model, modality settings may be specified by
the DR Calibration Tool.
If the DR Calibration Tool does not specify modality settings, refer to following
table.
Table 12: Modality settings
DR Table
DR Wall Stand
DX-D 300 U-arm
kV
75 kV
75 kV
75 kV
mAs (in-
dicative
value for
first ex-
posure
level)
28 mAs
32 mAs
32 mAs
Focus
large focus
large focus
large focus
SID
Use the (average) SID for normal operation.
Position
-
-
arm angle 0°
detector angle 0°
Grid
Remove the grid.
If the grid is fixed, use 32 mAs as indicative mAs-value.
Filtration
Mount the Agfa calibration filter (Cu 1.5 mm).
Check that there is no additional filtration.
DR Detector Calibration | Calibrating the DR Detector |
35
0134I EN 20210924 1204