DR 18M and DR 24M Exposure Sequence
Topics:
•
•
Modality Settings
The indicative exposure levels for each exposure are specified by the DR
Calibration Tool.
Table 6: Modality settings
kV
28 kV
mAs (indicative value for
first exposure level)
160 mAs
Filtration
MoMo
Note
:
If the exposure settings used for a previous calibration are
known, use these settings.
Exposure Sequence
Procedure:
1.
Allow the DR Detector to warm up for 30 minutes before performing the
calibration.
2.
Remove the compression paddle.
3.
Position the calibration PMMA block on the bucky.
4.
Set the mAs value to the indicative value for the exposure level.
5.
Make exposures as instructed by the DR Calibration Tool.
The rate of exposures is limited to one per minute.
24
| DR Detector Calibration | Calibrating the DR Detector