DR 10e C, DR 14e C and DR 17e C Exposure Sequence
Topics:
•
Using DR 10e C, DR 14e C and DR 17e C on multiple NX workstations
•
•
Using DR 10e C, DR 14e C and DR 17e C on multiple NX
workstations
If the DR Detector is used on multiple NX workstations, choose one NX
workstation that will be used for calibration and perform the regular
calibration each time on the same NX workstation. Do not calibrate the DR
Detector on another NX workstation.
Performing the calibration on multiple NX workstations leads to an
inconsistent situation, because the DR Detector is not able to store calibration
data for multiple NX workstations and each calibration completely discards
the calibration data of any previous calibration. Nevertheless each NX
workstation will independently request recalibration of the DR Detector every
12 months.
Modality Settings
Note
:
To calibrate DR 10e C, DR 14e C or DR 17e C, you have to
log in to a user account with administrator rights.
The indicative exposure levels for each exposure are specified by the DR
Calibration Tool.
Table 5: Modality settings
Focus
large focus
SID
Use the (average) SID for normal operation.
Position
If the DR Detector is used in a fixed position, use
the same position for calibration.
If the DR Detector is used in different positions,
perform the calibration with the short side of the
DR Detector parallel to the axis of the X-ray tube.
22
| DR Detector Calibration | Calibrating the DR Detector
0134I EN 20210924 1204